Multi-PID 2 photoionisation detectors from Draeger Safety Pacific



Multi-PID 2 photoionisation detector

Photoionisation detectors (PID) have featured in Draeger Safety’s product range for quite some time now. A while ago, the Dräger Multi-PID was presented.

Dräger Multi-PID 2: New device, new possibilities.

The Multi-PID 2 remains unchanged as far as its detector technology is concerned.

The Multi-PID 2 still offers high measurement sensitivity, accuracy, reproducibility of the measurement signal and low cross-sensitivity to water vapour (humidity).

The Multi-PID 1's familiar substance library and response factors have also been retained, meaning that no correction factors need to be taken into account when personnel switch to using the Multi-PID 2.

As regards handling and the instrument's display, however, the previous model has undergone some important changes:

Substance name displayed on screen:

The name of the substance currently calibrated is displayed in plain text on the screen, above the substance concentration.

If other substances occur at the site of measurement in addition to the one currently being detected, the total sum of organic substances is displayed, based on the calibrated substance.

Additional concentration information displayed on screen:

As well as the current concentration, the 15-minute average value (STEL), the 8-hour average value (TWL), the maximum recorded concentration (PEAK) and the elapsed measurement time are displayed on the screen in a smaller font size.

Data recording mode displayed on screen:

In the top right-hand section of the screen a symbol indicates whether the interval mode (i.e. continuous data recording) or the TAG mode (i.e. manual data recording) is activated, or whether no data are currently being recorded.

Just like its predecessor, the Multi-PID 2 can store 15,000 measured values. The Dräger GasVision then allows convenient processing and archiving of these data. GasVision outputs the maximum, minimum and average readings for each measurement interval.

One-key calibration:

Because PIDs generally require daily calibration, this process has been simplified in the Multi- PID 2 through the use of just one key. This means it is no longer necessary to work through several steps in the menu, so calibration could not be easier to perform.

Five languages in one device:

While users of the Multi-PID had to choose a language at the time of purchase, the Multi-PID 2 allows the operating language to be changed at will: German, English, French, Spanish and Italian are available, so personnel with different mother tongues can easily use the device too.

Applications:

PIDs offer multifunctional use when it comes to measuring organic and some inorganic substances.

The main difference as compared with other sensors (e.g. catalytic Ex-sensor or infrared Ex-sensor) is the fact that the PID is also able to conduct measurements right down to the sub-ppm range (with detection limits of approx. 0.5ppm). A number of typical applications are described below.

Leak detection and environmental analysis:

Due to its rapid response time and high level of sensitivity, a PID is well-suited to detecting leaks in many different settings, e.g. in pipelines in refineries, in tanks in chemicals factories, refineries and petrol stations and on ships and other means of transport.

If a leak is identified or an accident involving hazardous goods occurs, it is of course important to determine whether the soil in the vicinity of the leak has been contaminated.

Due to its high sensitivity, the Multi-PID 2 is also suited to such tasks, as well as to conducting a preliminary classification of soil samples (by way of a screening test to decide whether a laboratory analysis is necessary).

Arson investigation:

Arson investigations are often conducted to determine whether fire accelerants such as petrol, hydrocarbons or alcohol were used.

When police or fire officers inspect the scene of the fire, the Multi-PID 2 is used to search for any traces of such substances. In these situations the Multi-PID 2 is typically used with the sampling hose as a probe, allowing sites to be accessed which would otherwise be difficult to reach.

Entering aircraft tanks and other confined spaces:

Aircraft tanks constitute a particularly extreme example of confined spaces.

Since maintenance and inspection personnel have to conduct their checks right into the furthest corners of the tank, the use of respiratory protective equipment is only possible to a limited extent. This makes it all the more important to perform an all-clear measurement to ensure, for example, that the kerosene concentration is below 10ppm.

Because catalytic Ex- and infrared sensors cannot reliably measure such low concentrations of organic substances, this situation is also a typical example of a PID application.

In other tanks used to store organic substances too, an all-clear measurement should be conducted using a PID before personnel enter the tank. There is a clear trend towards assessing organic substances not only with respect to the explosion hazard they pose, but also with respect to their toxicity. This is something which can be done for a wide range of organic substances using the Multi-PID 2.

Fumigation Methyl bromide is still used, particularly when timber is transported overseas, but also during transport of textiles.

When unloading containers, or indeed when they are inspected by customs officials, it is important to ensure that concentrations of methyl bromide and any other fumigants are below the permissible limit values.

The advantage offered by the Multi-PID 2 here is that the instrument is able to cover an additional range of measurement.

Concentrations of between 0.5 and 2000ppm can be detected as standard, and the measurement range can be extended to 20,000ppm if the dilution probe is used. This makes the Multi-PID 2 suitable both for monitoring the efficiency of the fumigation process and for conducting measurements of relevance to health

Measuring diesel vapours:

Diesel is a mixture of different long-chain hydrocarbons and is used as fuel in cars, agricultural vehicles, trains and ships.

Up until now, attention was generally focused primarily on measuring concentrations of diesel fuels in the Ex concentration range (the lower explosive limit range, or LEL).

However, people's health may also be at risk below this concentration range. Bearing this in mind, the American Conference of Governmental Industrial Hygienists (ACGIH) issued a new limit value for the 8-hour mean value in 2002.

For the hydrocarbons in diesel, this limit is 100 mg/m3, equivalent to approx. 15ppm.

Diesel concentrations in the toxic concentration range are relevant when it comes to entering or cleaning tanks, and also during refilling and fuelling activities.

The Multi-PID 2 enables diesel concentrations in the ppm range to be determined. Because diesel is a mixture of different hydrocarbons, no universally applicable response factor can be stated.

For lighter diesel types the instrument can be calibrated to nonane and a response factor of 1.4 can be used. For heavier diesel types, such as those used in the shipping industry for example, a response factor of 1 can be applied.

Measurements in the chemicals and petrochemicals industry:

This is without doubt the main area of application for PIDs. Especially when personnel are to enter tanks, pipelines and other confined spaces, it is important to ensure that the concentration of the substances stored inside them is sufficiently low to prevent any harm being caused to the health of personnel.

The advantage of the PID, once again, lies in the fact that it allows measurements in the ppm range and that it can be calibrated to different substances.

Using the Multi- PID 2's calibration memory, different substance calibrations can be stored inside the instrument. This means that toluene can be measured first, for example, followed by an acetone measurement sometime later.

Calibration of the individual memories can be performed prior to the measurement process, so that only the calibration memory in question needs to be selected when it is time for measurement.

A further application in the chemicals and petrochemicals industry besides confined space measurements is of course classic workplace monitoring, for example when concentrations of hazardous substances need to be checked during particular work activities (e.g. filling of substances from one container to another).

In conclusion, it is clear that the Multi-PID 2 is able to perform measurements in many kinds of different applications and work areas.

Through Draeger Safety's technical applications service, users have access to far more response factors than those already stored in the substance library of the Multi-PID 2. For substances not listed there it is possible to determine whether they can be measured by means of PID.

The Multi-PID 2 photoionisation detectors are available from Draeger Safety Pacific.

15-Oct-2007

Contact Draeger Safety Pacific

Name:
Company Name:
Contact Number:
Email Address:
Your State:
Your Message:

Please uncheck this box if you don't want to receive the latest product updates in Ferret Daily Product News, Ferret's daily eNewsletter.

does not match
 
Your contact details will be passed onto this company ( see our Privacy Policy )
More Articles

Draeger Safety Pacific News


Dräger X-am 3000 mobile multi-gas detector New version mobile multi-gas detector available from Draeger Safety Pacific (16-Oct-2007)
Dräger X-am 3000 is a mobile multi-gas detector for the continuous measurement of hydrogen sulphide, carbon monoxide, oxygen and explosive gases.

Alcotest 6510 breath alcohol analyser New breath alcohol analyser available from Draeger Safety Pacific (12-Oct-2007)
The breath alcohol analysers have low cross-sensitivities with other non-alcoholic substances which may be present in the exhaled air.

Dräger X-am 5000 gas detection instrument New version gas detector available from Draeger Safety Pacific (9-Oct-2007)
The Dräger X-am 5000 is tough: water and dust-resistant according to IP 67, the detector remains fully functional even after falling into water.

Mobile training vehicle D•pod Drager’s mobile training vehicle at Sydney Safety Show (5-Oct-2007)
Drager’s mobile training vehicle, the Drager Personnel Occupational Development trainer (D•pod) is designed to permit delivery of the required occupational training modules associated with workplace safety training.

News Sign up View all  |  RSS Feed

Related Articles

PST-T0110 temperature probes, transducers and transmitters available from Pacific Sensor Technologies PST-T0110 temperature probes, transducers and transmitters available from Pacific Sensor Technologies (3-Dec-2008)
Pacific Sensor Technologies offer the PST-T0110 programmable temperature transmitter that is suitable for Pt1000/3850ppm with 4-20 mA output.

Madgetech’s QuadProcess DC current recorder available from Pacific Sensor Technologies Madgetech’s QuadProcess DC current recorder available from Pacific Sensor Technologies (2-Dec-2008)
Pacific Sensor Technologies offer the QuadProcess, a 4-channel, low level, -20mA to 100mA DC current recorder, manufactured by Madgetech.

Harting’s crimp version HC 350 high voltage contacts available from Soanar Harting’s crimp version HC 350 high voltage contacts available from Soanar (2-Dec-2008)
HARTING, distributed in Australia and New Zealand by Soanar - Electronic Component & Supply Chain Solutions, have extended the Han Connector product range, the HC 350 contact, which comes with a rating up to 350 Amps, 2Kv and 4Kv with an adaptor.

Pacific Sensor Technologies release PST-C370 RTD water proof thermometer Pacific Sensor Technologies release PST-C370 RTD water proof thermometer (1-Dec-2008)
Pacific Sensor Technologies have released a new, high accuracy, water proof thermometer, suitable for food, industrial and laboratory environments.

Pacific Sensor Technologies unveil Level1000 temperature and water level data recorder Pacific Sensor Technologies unveil Level1000 temperature and water level data recorder (1-Dec-2008)
The Level1000, from Pacific Sensor Technologies, is a rugged, submersible, temperature and water level data recorder. Pacific Sensor Technologies’ Level1000 level data logger has a stainless steel enclosure thta makes it suitable for use in harsh, chemical and corrosive environments.

Access over 2000 Manufacturing and Operations jobs online!